The OdaLog has been specifically designed for
the Wastewater Industry, primarily for locating at or close
to the source of Hydrogen Sulphide emissions such as within
sewerage pumping stations and receiving manholes.
Its purpose is to log levels of H2S in the
range 0-200 ppm for later downloading of data in tabular
or graphical form to highlight significant variations in
H2S levels over time.
Due to the nature of the environment in which
the instrument has to operate, it has been designed to be
splash resistant, robust, small, and able to log data for
extended periods of time.
The following is detailed information on this
globally recognised product.
Ability to customise graphical scale and
mark time intervals on X-axis
Records OdaLog L2 serial number on printout
and allows graphs to be labelled
Uses Infra Red Communication for downloading
and configuring
42,000 logging intervals allow very detailed
or very long studies e.g.
Over 4 months at 5 minute intervals
Over 4 weeks at 1 minute intervals
Over 4 days at 10 second intervals
Instrument Temp Range
-20°C (-4oF) to 50°C (122oF)
Sensor Temp Range
-10°C to +40°C (14oF to 104oF)
Instrument Relative Humidity Range
15-90% non-condensing
Ingress Protection
IP66/68
External Dimensions
Diameter 62mm (2.44”) x Height 196mm (7.72”)
Instrument Weight
420g (14.8oz)
Power Supply(s)
Replaceable C-Size Lithium Battery (see approved battery type/brand/make only)
Battery Life/ Run Time
Greater than 8months (for life of Lithium battery)
Data Logging Capacity
Up to 42,000 readings in total
Interval
Selectable from 1 second to 1 hour
Data Logging Duration Example
Capacity > Logging Interval of 1min = 29 days
Capacity > Logging Interval of 5min = 6+ months
The following is a list of gas types and ranges available for the OdaLog® L2
Model
Gas
Range
Resolution
Accuracy
(Full Scale)
Accuracy (@ 20ppm gas, STP)
Long term Logger
(LL)
H2S
Hydrogen Sulphide
0-1000 ppm
1 ppm
1% Full Scale
± 2ppm
0-200 ppm
0.1 ppm
1% Full Scale
Standard
Logger
(SL)
0-2000 ppm
1 ppm
2% Full Scale
0-1000 ppm
1 ppm
1% Full Scale
0-200 ppm
0.1 ppm
1% Full Scale
0-50 ppm
0.1 ppm
4% Full Scale
SL
CO
Carbon Monoxide
0-1000 ppm
1 ppm
2% Full Scale
SL
SO2
Sulphur Dioxide
0-20 ppm
0.5 ppm
10% Full Scale
SL
NO
Nitric Oxide
0-100 ppm
1 ppm
2% Full Scale
SL
NO2
Nitrogen Dioxide
0-50 ppm
0.1ppm
4% Full Scale
SL
CL2
Chlorine
0-20 ppm
0.1 ppm
10% Full Scale
SL
HCN
Hydrogen Cyanide
0-30 ppm
0.2 ppm
6% Full Scale
SL
NH3
Ammonia
0-100 ppm
1 ppm
10% Full Scale
NOTE: STP = Standard Temperature and Pressure:20 °C (68 °F) @ absolute pressure 101.3 kPa (14.7 psi).
Authority
Certificate Number
Code
ITACS
IECEx ITA 08.0010
Ex ia IIC T4 (TAMB = -20oC to +50oC) IP66/68
NEMKO
Nemko08ATEX1270
II 2 G,Ex ia IIC T4 IP66/68
NOTE:
Click on IECEx and ATEX certification for further information
We recommend that the OdaLog is returned to an authorised
App-Tek service centre at least once every six months for a full inspection,
software upgrade as developed, calibration, test for linearity and
to have new batteries fitted.
12 month warranty when used in accordance with the operator's
manual (excluding calibration costs).
In the interest of continued improvement, we reserve
the right to change design features without prior notice.